HVLD Test Parameters, Test Methods 1 through 6
Test Method | Clip (none, up, down) | Test direction | High voltage set point (kV) | Sensitivity set point lower limit | Sensitivity set point upper limit |
---|---|---|---|---|---|
1 | None | Septum first | 12.5 | 275 | 285 |
2 | None | Bag first | 12.5 | 255 | 265 |
3 | Up | Septum first | 12.5 | 250 | 260 |
4 | Up | Bag first | 12.5 | 240 | 250 |
5 | Down | Septum first | 12.5 | 255 | 265 |
6 | Down | Bag first | 12.5 | 240 | 250 |