Typical Operating Parameters, LC/UV/MS Analyses
Operating Parameter | Operating value | |
---|---|---|
Column | Agilent Zorbax Eclipse Plus C18, 100 × 3.0 mm, 3.5 μm particles | |
Column Temperature | 40–50 °C | |
Mobile Phase Components | A = 10 mM ammonium acetate, B = acetonitrile | |
Mobile Phase Gradient | Time (min) | % B |
0.0 | 5.0 | |
8.4 | 100.0 | |
35.0 | 100.0 | |
36.0 | 5.0 | |
39.0 | 5.0 | |
Mobile Stage Flow Rate | 0.8 mL/min | |
Injection Size | 10–50 μL | |
Detection, UV | 205–300 nm; spectra recorded at λ = 210, 220, 230, 250, and 270 nm; 205 nm for variable wavelength detector | |
Detection, MS | API-ES or APCI, positive ion and negative ion (mass range 80–1500) | |
Instrumentation Used |
| |
Sample Preparation | None, direct injection |