Total Particles, Viable Particles, and MCPs per 141.5 L (5 ft3) or per 15 s in the Unidirectional Airflow Workstation Air
Test | BioTrak® | Lasair II® | AirTrace® | Gelatine Filter | |||
---|---|---|---|---|---|---|---|
Viables | ≥0.5 μm | ≥5 μm | ≥0.5 μm | ≥ 5μm | |||
Air Sampler Counts per 141.5 L (5 ft3) | |||||||
1. At rest | 0 | 1 | 0 | 0 | 0 | 0 | 0 |
2. Set-up | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
3. Production | 0 | 0 | 0 | 2 | 1 | 0 | 0 |
Air Sampler Counts per 15 s | |||||||
4. Incident 1 (sterilised forceps) | 0 | 1 | 0 | 2 | 1 | 0 | 0 |
5.Incident 2 (gloved hands) | 0 | 1 | 0 | 0 | 0 | 0 | 0 |
6.Incident 3 (ungloved hands) | 2 | 87 | 11 | 71 | 8 | 0 | 0 |