Table VI

Total Particles, Viable Particles, and MCPs per 141.5 L (5 ft3) or per 15 s in the Unidirectional Airflow Workstation Air

TestBioTrak®Lasair II®AirTrace®Gelatine Filter
Viables≥0.5 μm≥5 μm≥0.5 μm≥ 5μm
Air Sampler Counts per 141.5 L (5 ft3)
1. At rest0100000
2. Set-up0000000
3. Production0002100
Air Sampler Counts per 15 s
4. Incident 1 (sterilised forceps)0102100
5.Incident 2 (gloved hands)0100000
6.Incident 3 (ungloved hands)2871171800