Table III

Examples of FTIR Silicone Assay Data Obtained for Multiple Lots of Syringes with Ages of Less than 4 Months with 0.25 and 0.4 mg Nominal Initial Silicone Oil Targets

LotInitial Nominal Target (mg)Silicone Assayed by FTIR (μg/syringe ± SD)Percent of Initial Nominal Target (%)Coefficient of Variation, CV (%)
A0.25210 ± 16 (n = 10)848
B0.25193 ± 12 (n = 10)778
C0.4436 ± 29 (n = 10)1097
D0.4358 ± 35 (n = 10)9010
E0.4357 ± 45 (n = 10)8913
F0.4361 ± 15 (n = 10)904