UHPLC/UV/MS Operating Conditions
Parameter | Value | ||
---|---|---|---|
Column: | Waters Aquity UHPLC BEH C18, 3.0 mm × 100 mm, 1.7 μm | ||
Column Temperature: | 60 °C | ||
Column Flow Rate: | 0.7 mL/Minute | ||
Mobile Phase A: | Water | ||
Mobile Phase B: | Methanol | ||
Strong Wash: | Methanol | ||
Weak Wash: | 80/20 Water/Methanol | ||
Seal Wash: | 90/10 Water/Methanol | ||
Gradient Program: | Time (min) | % Mobile Phase B | Curve |
0 | 20 | NA | |
12 | 95 | 6 | |
15 | 95 | 6 | |
15.5 | 100 | 6 | |
20 | 100 | 6 | |
22 | 20 | 1 | |
Injection Volume: | 5 μL | ||
Injection Mode: | Partial Loop with Needle Overfill | ||
Strong Wash Volume: | 600 μL | ||
Weak Wash Volume: | 1800 μL | ||
Loop Size: | 10 μL | ||
PDA Detector: | 220–400 nm |
MS Operating Conditions | ||
---|---|---|
Parameter | APCI+ | APCI− |
Scan Range (Daltons): | 110–1250 | 130–1250 |
Corona Current (μA): | 0.5–2 | 2–5 |
Cone Voltage (V): | 20–35 | 50 |
Extractor (V): | 3.0 | 3.0 |
RF (V): | 0.1 | 0.1 |
Source Temperature (°C): | 120 | 120 |
Probe Temperature (°C): | 550 | 550 |
Cone Gas Flow (L/h): | 50 | 50 |
Desolvation Gas Flow (L/h): | 600 | 600 |