Table I

UHPLC/UV/MS Operating Conditions

ParameterValue
Column:Waters Aquity UHPLC BEH C18, 3.0 mm × 100 mm, 1.7 μm
Column Temperature:60 °C
Column Flow Rate:0.7 mL/Minute
Mobile Phase A:Water
Mobile Phase B:Methanol
Strong Wash:Methanol
Weak Wash:80/20 Water/Methanol
Seal Wash:90/10 Water/Methanol
Gradient Program:Time (min)% Mobile Phase BCurve
020NA
12956
15956
15.51006
201006
22201
Injection Volume:5 μL
Injection Mode:Partial Loop with Needle Overfill
Strong Wash Volume:600 μL
Weak Wash Volume:1800 μL
Loop Size:10 μL
PDA Detector:220–400 nm
MS Operating Conditions
ParameterAPCI+APCI−
Scan Range (Daltons):110–1250130–1250
Corona Current (μA):0.5–22–5
Cone Voltage (V):20–3550
Extractor (V):3.03.0
RF (V):0.10.1
Source Temperature (°C):120120
Probe Temperature (°C):550550
Cone Gas Flow (L/h):5050
Desolvation Gas Flow (L/h):600600