Table XVIII

Summary of Reported Occurrences for the ICH Q3D Class 1 Metals (As, Cd, Hg, Pb) in Material Extracts

Type of TestConcentration (mg/kg unless noted)Extraction ConditionsMaterial and Sample TypeRef.
SolventConditions
Arsenic (As)
Extraction0.55 ng/mLWaterGlass syringe, 70 °C for 96 hMean result for multiple glass syringes25
0.01pH 2.5Sealed vessel, 121 °C for 1 h, 0.025 g/mL solventPlasticized PVC resin35
<0.01pH 9.5
1–38 ng/mLEthanolFilled portions of silicone tubing, 55 °C for 24 h6 commercially available silicone materials54
Cadmium (Cd): No reported values
Lead (Pb)
ExtractionNR1–4.1Low pH (≈3)Sealed vessel, 55 °C for 4 days, 0.1–1.0 g sample/mL solvent5 glass materials tested; vials and parts12
NR–2.6High pH (≈9)
NREthanol/water
60 ng/mL0.1% EDTAFilled vials, 60 °C for 23 weeksType 1 clear glass vials23
20 ng/mLWater
40 ng/mLpH 6.8 buffer
40 ng/mL0.1% EDTAFilled vials, 60 °C for 23 weeksType 1 amber glass vials23
30 ng/mLpH 6.8 buffer
20 ng/mL0.1% EDTAFilled vials, 60 °C for 23 weeksType 1 SiO2-coated glass vial23
10 ng/mLWater
10 ng/mLpH 6.8 buffer
10 ng/mLWaterSealed vessel, 60 °C for 23 weeks, 0.25 cm2/mL solventLow-density polyethylene resin23
10 ng/mLpH 6.8 buffer
10 ng/mL0.2% Polysorbate 80
10 ng/mLWaterSealed vessel, 60 °C for 23 weeks, 0.25 cm2/mL solventHigh-density polyethylene resin23
30 ng/mLpH 6.8 buffer
30 ng/mL0.2% Polysorbate 80
1.6 mg/LAqueous buffersCut strips extracted at 60 °C for 10 days, 1 cm2/2 mL solventPlasticized PVC blood transfusion bag16
0.01pH 2.5Sealed vessel, 121 °C for 1 h, 0.025 g/mL solventPlasticized PVC resin35
<0.01pH 9.5
1.4pH 2.5Various extraction methods, 0.02–0.04 g/mL solventPlasticized PVC resin56
30 ng/mL0.1% EDTA40 °C for 23 weeks, 10 mL solvent in 75 mL bottlePolyethylene terephthalate bottle23
Mercury (Hg): No reported values
  • Note:

  • 1 NR = Not reported: not extracted in reportable quantities, generally less than 0.05 mg/L or 0.1 mg/kg.