Table III

Fluorescence Measurements* Using Optimized Instrument Parameters

Measurement NumberMethylene Blue in SyringeRhodamine in Syringe
Concentration 0.2 ug/mLConcentration 0.2 ug/mLConcentration 0.05 ug/mLConcentration 0.005 ug/mLConcentration 0.005 ug/mL
1 Syringe, 10 Readings, Re-positioning the Syringe for Each Reading10 Syringes, 1 Reading for Each1 Syringe, 10 Readings, Re-positioning the Syringe for Each Reading1 Syringe, 10 Readings, Re-positioning the Syringe for Each Reading10 Syringes, 1 Reading for Each
189.694.8964.2105.5144.6
287.095.5970.0107.1138.8
388.589.2964.2106.6138.8
488.3107.4964.2105.7155.7
590.5115.5986.1107.7148.8
688.079.1964.8105.4141.9
789.2106.5959.1105.8149.0
887.097.5964.2106.5120.8
986.8114.3957.0109.0146.6
1088.0108.5958.3105.8149.6
Mean88.3100.8965.2106.5143.5
SD1.211.68.31.19.5
% RSD1.411.50.91.16.6