Table IV

Comparison of D-Values of Test BI in Water for Injection Measured in Different Heating Apparatus (BIER and Oil Bath)

MeasurementD105-Value
BIEROil Bath
10.921.00
20.911.08
30.951.03
Mean0.931.04
SD0.010.03
  • A glass capillary and 0.1 mL of test solution were used. Test BI is Bacillus oleronius.