Table VI

Results of Measurements on Certified Standards of Known Carbon Dioxide Gas Pressure in Units of torr throughout the 14 Day Study

Carbon Dioxide Standard Value (torr)Measured Carbon Dioxide Pressure (torr)
AverageStandard Deviation (SD)
49.350.10.2
99.3100.20.4
149.2149.10.6
544.4542.62.1