Table II

Leak Testing by He-Leakage Test Method (HLT-560, Pfeiffer Vacuum AG, Zürich, Switzerland) on Empty 2 mL Vials Crimped at 40 N (n = 10)

Cu wire diameter (μm)Equivalent diameter of a circular-shaped aperture (μm) when applying 30 N capping forceEquivalent diameter of a circular-shaped aperture (μm) when applying 50 N capping forceNumber of leaking vials*Leak rate—order of magnitude (mbar × L/s)
0000/10<10−8
151290/10<10−8
2016132/1010−5
2822186/1010−4
40312510/1010−4
60473810/1010−3
  • * Acceptance criteria leak rate <1 × 10–7 mbar × L/s (tight system).