Accuracy (%) of Conductivity Measurements
Purified Water with Conductivity Standard Spike at: | ||||||
---|---|---|---|---|---|---|
0 μS/cm | 1 μS/cm | 5 μS/cm | 10 μS/cm | 25 μS/cm | 100 μS/cm | |
Conductivity (μS/cm) | 0.68 | 1.70 | 5.42 | 10.18 | 24.05 | 99.30 |
Accuracy (%)* | – | 102 | 95 | 95 | 93 | 99 |
↵* Accuracy (%) = [(Conductivity of Spiked – Conductivity of Blank Water)/Theoretical Conductivity] × 100.