TABLE VI

Correlation between Conductivity (μS/cm) and Titration Volumes (MilliLiters of 0.01 M HCl added to 25 mL Vial Extract) of Water Extracts of 3 cc Glass Vials

SetAnalysisAnalyses: A, Conductivity (μS/cm); B, Titration (mL)Correlation
1234567SlopeInterceptR2
1A10.164.027.011.8710.085.364.450.0302−0.00980.9902
B0.300.100.200.060.300.140.13
2A9.853.727.711.8610.415.424.590.02720.00650.9978
B0.280.110.210.060.290.150.13
3A9.904.207.502.0210.205.664.720.0296−0.00860.9726
B0.270.110.200.060.320.150.14
4A10.123.487.072.079.915.724.520.0290−0.00490.9904
B0.300.090.200.060.270.170.12
5A10.153.337.142.119.355.633.880.0285−0.00230.9934
B0.280.090.200.060.270.170.10
6A10.793.537.432.099.945.794.620.02620.01030.9910
B0.290.100.200.060.270.180.13
7A10.153.797.521.8510.055.755.640.0272−0.00810.9922
B0.280.090.190.050.260.150.14
8A10.323.537.352.1210.825.934.930.0275−0.00840.9797
B0.300.090.190.050.270.150.13
9A10.533.497.422.0310.635.414.900.02530.00380.9937
B0.280.090.180.060.270.140.13
Average0.0279−0.00240.9890
RSD (%)6n/a1