TABLE III

Typical Operating Parameters, GC/MS Analyses

Operating ParameterOperating Value
ColumnRestek (Bellefonte, PA) Rtx®-VMS, 30 m × 0.25 mm, 1.4 μm film thickness
Oven ProgramStart at 40 °C, hold for 4 min; ramp at 25 °C/min to 220 °C; ramp at 5 °C/min to 250 °C; ramp at 25 °C/min to 260 °C, hold for 8 min
Carrier GasHe at 1.8 mL/min
InjectionSplitless; 0.25 μL
Injector Temperature280 °C
MS Transfer Line Temperature260 °C
MS Detection DetailsEI (70.3 ev), mass range 29–550 amu
Instrumentation UsedAgilent (Santa Clara, CA) 7890 GC/7683B autosampler/7000B Triple Quadrapole MS (Agilent Mass Hunter B.07.01 data system)