TABLE II

Operating Parameters, GC/FID/MS Analysis of the Static Extracts

Operating ParameterOperating Value
ColumnJ&W DB-5HT, 30 m × 0.25 mm × 0.1 μm
Oven ProgramStart at 40 °C (1 min hold). Ramp at 10 °C/min to 280 °C, hold for 2 min. Ramp at 15 °C/min to 310 °C, hold for 5 min
Carrier GasHe at 2 mL/min
Injection Volume and Mode2 μL; splitless for 0.5 min
Injector Temperature300 °C
FID Detector Temperature325 °C
MS Transfer Line Temp310 °C
MS Detection Details70 eV (+), mass range of 29–800 amu
Solvent Delay3–5 minu for non-derivatized samples; 6–8 min for derivatized samples