TABLE III

Particle Count Analysis by Coulter Counter of mAb A Solution after Mixing at 200 rpm for up to 1 h in a Mavag Mavadrive™ Mixer with Different Impeller Designs

Particle Count (particle/mL)a
Mixing Time (h)Impeller with Blades (2.5 L of mAb A)Impeller without Blades (2.5 L of mAb A)Impeller without Blades but with Attached Plastic Cylinder Dividerb
Outer Reservoir (2.5 L of mAb A)Inner Reservoir (10 mL of mAb A)c
02.55 × 104 ± 5.24 × 1031.45 × 104 ± 4.08 × 1031.24 × 104 ± 8.55 × 1021.88 × 104 ± 1.01 × 104
0.51.74 × 105 ± 6.68 × 1041.93 × 105 ± 1.46 × 1041.94 × 104 ± 9.80 × 1031.86 × 107 ± 2.28 × 106
1.04.67 × 105 ± 8.71 × 1045.03 × 105 ± 2.77 × 1041.29 × 104 ± 3.02 × 1032.77 × 107 ± 5.29 × 106 d
Change in SvP Relative to t = 0e183511471
  • a Particle count data represents an average of two separate trials.

  • b Particle count in the outer reservoir is an average of counts in samples taken from areas close to the impeller and away from the impeller.

  • c Particle count before volume correction. Particle count after volume correction can be measured by diluting a sample from the inner reservoir 250 times using the solution in the outer reservoir.

  • d The average particle count of the diluted material was determined to be 2.60 × 105 particles/mL.

  • e Change in SvP relative to t = 0: average particle count after mixing for 1 h ÷ average particle count in mAb A at t = 0.