TABLE IV

Two Examples of Potential LAL False Positives in Biopharmaceutical Drug Substance Manufacturing and the Insensitivity of rFC toward Suspected Beta Glucans from Cellulosic Filter Material

SampleSite 1 LAL Result (EU/mL)Site 2 LAL Result (EU/mL)rFC (EU/mL)GB Buffer +LAL (EU/mL)PTS™ (beta glucan pg/mL)
10.1410.0946<0.08<0.042265
20.120.0712<0.04<0.041254
30.1050.0748<0.02<0.04970
40.1320.157<0.02<0.083173
50.1760.109<0.04<0.082798
LRW Negative ControlBelow detection limit<10
10.7370.187<0.100<0.5001379
20.6810.171<0.100<0.5001382
30.6690.134<0.100<0.5001076
  • LRW, LAL reagent water; GB Buffer, glucan blocking buffer.