Operating Parameters, LC/MS/UV Analysis of the Static Extracts
Operating Parameter | Operating Value | |
---|---|---|
Column | Agilent ZORBAX Eclipse Plus C18, 2.1 × 100 mm, 3.5 um particle size | |
Column Temperature | 40°C | |
Mobile Phase Components | A = 10 mM Ammonium acetate, B = methanol | |
Mobile Phase Gradient | Time (min) | % B |
0.0 | 5 | |
1.0 | 5 | |
20.0 | 99 | |
30.0 | 99 | |
30.1 | 5 | |
35.0 | 5 | |
Mobile Phase Flow Rate | 0.5 mL/min | |
Sample Size | 60 μL | |
Detection, UV | 210, 230, 250, 280 nm | |
Detection, MS | API-ES, positive ion and negative ion (range 80–1200) |