TABLE IV

Recovery of Protein A and Protein A Dimer from Stored Prefillable Syringes after 1 Year

SampleSoluble Protein Recovered (SE-HPLC)Dimer (Percent of Soluble Protein)
Control100.00 ± 0.272.55 ± 0.15
Glass—5 °C99.73 ± 0.612.60 ± 0.15
CZA—5 °C100.87 ± 0.272.46 ± 0.08
CZE—5 °C100.54 ± 0.172.51 ± 0.07
Glass—25 °C99.69 ± 0.624.90 ± 0.33
CZA—25 °C101.55 ± 0.662.85 ± 0.16
CZE—25 °C101.51 ± 0.822.75 ± 0.12