TABLE V

Summary of Parameters Associated with the Model Sample Vials Sealed with a Crimped Stopper-Cap Assembly and Containing a Lenox Laser-Drilled Micro-Hole in the Glass Wall

Sample VialDefectDiffusive FlowEffusive Flow
Measured Diameter (μm)Idealized Area (μm2)Diffusion Parameter (cm3/s)RSQEffusion Parameter (cm3/s)RSQ
L2-12.13.373.25E-050.999394.70E-040.99929
L2-22.23.916.78E-050.997629.54E-040.99905
L2-32.34.237.28E-050.997741.41E-030.99939
L2-42.03.245.85E-050.999646.63E-040.99771
L2-52.44.347.76E-050.995131.44E-030.99939
L5-15.221.167.49E-050.998472.84E-030.99972
L5-25.221.161.24E-040.999013.78E-030.99975
L5-35.624.373.38E-050.999054.82E-030.99988
L5-45.019.406.97E-050.999583.28E-030.99986
L10-19.368.221.28E-040.999409.48E-030.99986
L10-210.079.013.02E-040.995231.36E-020.99955
L10-39.469.841.57E-040.998711.18E-020.99984
L10-410.077.911.89E-040.997041.46E-020.99988
L15-114.9173.201.62E-040.998953.32E-020.99846
L15-215.3183.378.54E-050.998883.06E-020.99911
L15-313.7147.413.75E-040.998222.71E-020.99981
L15-414.3161.282.14E-040.997934.30E-020.99966
L15-515.7192.611.88E-040.998833.98E-020.99952
  • The coefficient of determination for the least-squares fit associated with the calculation of the corresponding rate.