Comparison of WLI, 3D-LSM, and Combined WLI+LAI for the Characterization of the Silicone Oil Layer Thickness in Empty Glass Syringes
WLI | 3D-LSM | WLI+LAI | |
---|---|---|---|
Thickness range | 80–4000 nm | >10 nm | 20–4000 nm |
Layer imaging | − | + | − |
Single measurement time | ∼4 s | ∼60 s | ∼15 s |
Radial thickness mapping | ++ | + | ++ |
Longitudinal thickness mapping | ++ | + | ++ |
Sample preparation | + | − | + |