Comparison of CCI Results Obtained by Headspace Analysis (Performed at Room Temperature, After 1 Day of Storage in Dry Ice) and by the Helium-Leakage Test (ThermCCI) Measured Directly at −60°C (After 1 Day of Storage in Dry Ice) in 2 mL Vials with Serum Stoppers
Storage Conditions: Static | CO2 Headspace (Number of Leaking Vials/Total Vials) | Helium-Leak @ −60°C (Number of Leaking Vials/Total Vials) |
---|---|---|
RTa | 0/10 | 2/10 |
−75°C (*)b | 0/10 | 3/10 |
Storage Conditions: Dynamic | CO2 Headspace (Number of Leaking Vials/Total Vials) | Helium-Leak @ −60°C (Number of Leaking Vials/Total Vials) |
---|---|---|
RT | 0/10 | 2/10 |
−75°C (*) | 0/10 | 2/10 |