Temperature Programs (°C) for the Measurement of Ba, Cd, Pb, Sn, and Zn for the Different Samples Analyzed
Element | Matrix | Modifier | Pyrolysis | Atomization |
---|---|---|---|---|
Ba | Bags | Yes | 1000 | 2250 |
NaCl/Glucose/Tween 80 | Yes | 1150 | 2600 | |
Cd | Bags/NaCl/Glucose | No | 600 | 1100 |
Tween 80 | Yes | 850 | 1350 | |
Pb | Bags/NaCl/Glucose/ | No | 700 | 1500 |
Tween 80 | Yes | 850 | 1600 | |
Sn | Bags/Tween 80 | No | 800 | 2400 |
NaCl | Yes | 1150 | 2450 | |
Glucose | Yes | 1400 | 2350 | |
Zn | Bags | No | 500 | 1250 |
NaCl/Glucose | No | 600 | 1250 | |
Tween 80 | Yes | 900 | 1550 |