TABLE III

Results of Test Method B Applied on Syringes, Cartridges, and Vials Expressed in parts per milliona

Container TypeAl2O3 (ppm)B2O3 (ppm)SiO2 (ppm)
MeanStd. Dev.MeanStd. Dev.MeanStd. Dev.
1 mL Long Syringes
    Bulk1.100.102.060.1514.921.34
    Siliconized1.260.152.320.2716.651.58
    Alba<LoQbc0.660.10<LoQb
    LoQb0.380.654.28
3 mL Vials
    Bulk1.750.243.300.7117.242.23
    Alba0.270.050.650.124.180.51
    LoQb0.190.322.14
3 mL Cartridges
    Bulk0.920.101.870.2113.061.54
    Alba0.230.050.590.083.770.26
    LoQb0.190.322.14
  • a For each category 50 containers were analyzed. “Bulk” refers to untreated samples, “Siliconized” refers to samples treated with silicone oil process, and “Alba” refers to samples siliconized with Alba process.

  • b LoQ = Limit of Quantification.