Number of Tested Samples that have Shown Bacterial Ingress at a Certain Defect Size Compared to the Total Number of Tested Samples for PE and EVA Films at 300 mbar Pressure
Defect Size (µm) | Bacterial Ingress / Total Samples | |
---|---|---|
PE Film (400 µm thick) | EVA Film (300 µm thick) | |
1 | 0/30 | 0/30 |
2 | 2/30 | 0/30 |
3 | 2/30 | 9/30 |
5 | 10/30 | 17/30 |
10 | 22/30 | 15/30 |
Total Positives | 36/150 | 41/150 |