TABLE III

Number of Tested Samples that have Shown Bacterial Ingress at a Certain Defect Size Compared to the Total Number of Tested Samples for PE and EVA Films at 300 mbar Pressure

Defect Size (µm)Bacterial Ingress / Total Samples
PE Film (400 µm thick)EVA Film (300 µm thick)
10/300/30
22/300/30
32/309/30
510/3017/30
1022/3015/30
Total Positives36/15041/150