Sensitivity and specificity of the test kit BAX for screening/E. coli O157:H7 in ground beef: independent laboratory study

J AOAC Int. 2000 Nov-Dec;83(6):1349-56.

Abstract

An independent laboratory study of the BAX for Screening/E. coli O157:H7 kit was conducted at the National Food Laboratory, Inc., Dublin, CA, to complete AOAC Performance Tested Method certification. The BAX system kit was compared with the BAM culture method and a modified BAM culture method for detection of E. coli O157:H7 in ground beef. The BAX system kit detected the target organism at levels approximately 10-fold lower than those that gave positive BAM results. This study validated product claims, and Performance Tested Method status was granted.

Publication types

  • Validation Study

MeSH terms

  • Animals
  • Cattle
  • Escherichia coli O157 / chemistry*
  • Freezing
  • Meat / microbiology*
  • Reagent Kits, Diagnostic

Substances

  • Reagent Kits, Diagnostic