“Durapore,” “Sartobran,” and “Ultipore Nylon 66” membrane filter cartridges, rated by the manufacturers at 0.2-μm poresize, removed all Pseudomonas diminuta ATCC 19146 from saline lactose broth and all bacteria and molds from naturally-contaminated trypticase soy broth. After autoclaving of the Durapore and Sartobran cartridges, their bubble points increased by approximately 5 and 15 psi, respectively, whereas the Ultipore Nylon 66 cartridge bubble points did not change appreciably. Because the pre-autoclave bubble points of the Durapore and Sartobran made a satisfactory diffusion test at 80% of the post-autoclave bubble point difficult or impossible to do, these were best diffusion tested at 30 psi. However, the Ultipore Nylon 66 could be diffusion-tested at 40 psi. The diffusion test proved to be an unambiguous measure of filter integrity, but the bubble point test also was deemed essential to the proper characterization of 0.2-μm poresize membrane filter cartridges. The Ultipore Nylon 66 cartridges, rated by the manufacturer at 0.45 μm, tolerated pressures in excess of 50 psi in the filtration of large volumes parenterals. These cartridges were diffusion tested at 25 psi. Bubble point analysis was considered less critical for the filters used with large volume parenterals than for the 0.2-μm poresize filters.
- Received February 13, 1981.
- Accepted May 15, 1981.
- Copyright © Parenteral Drug Association. All rights reserved.
PDA members receive access to all articles published in the current year and previous volume year. Institutional subscribers received access to all content. Log in below to receive access to this article if you are either of these.
If you are neither or you are a PDA member trying to access an article outside of your membership license, then you must purchase access to this article (below). If you do not have a username or password for JPST, you will be required to create an account prior to purchasing.
Full issue PDFs are for PDA members only.
Note to pda.org users
The PDA and PDA bookstore websites (www.pda.org and www.pda.org/bookstore) are separate websites from the PDA JPST website. When you first join PDA, your initial UserID and Password are sent to HighWirePress to create your PDA JPST account. Subsequent UserrID and Password changes required at the PDA websites will not pass on to PDA JPST and vice versa. If you forget your PDA JPST UserID and/or Password, you can request help to retrieve UserID and reset Password below.