Abstract
Spores of Geobacillus stearothermophilus are very sensitive to changes in temperature. When validating sterilizing processes, the most common bioindicator (BI) is spores of Geobacillus stearothermophilus ATCC12980 and ATCC7953 with about 106 spores /BI and a D121-value of about 2 minutes in water. Because these spores of Geobacillus stearothermophilus do not survive at a F0-value above 12 minutes, it has not been possible to evaluate the agreement between the biological F-value (FBIO) and physical measurements (time and temperature) when the physical F0-value exceeds that limit.
However, it has been proven that glycerin substantially increases the heat resistance of the spores, and it is possible to utilize that property when manufacturing BIs suitable to use in processes with longer sterilization time or high temperature (above 121°C). By the method described, it is possible to make use of the sensitivity and durability of Geobacillus stearothermophilus' spores when glycerin has increased both test range and accuracy.
Experience from years of development and validation work with the use of the highly sensitive glycerin-water-sporesuspension sensor (GWS-sensor) is reported. Validation of the steam sterilization process at high temperature has been possible with the use of GWS-sensors. It has also been shown that the spores in suspension keep their characteristics for a period of 19 months when stored cold (8°C).
- Bioindicator
- Geobacillus stearothermophilus
- GWS-sensor
- Steam sterilization
- Sterilization safety
- F0-concept
- FBIO-value
- F-value
- Category of submission: Validation
- Technology
Footnotes
- Copyright © Parenteral Drug Association. All rights reserved.
PDA members receive access to all articles published in the current year and previous volume year. Institutional subscribers received access to all content. Log in below to receive access to this article if you are either of these.
If you are neither or you are a PDA member trying to access an article outside of your membership license, then you must purchase access to this article (below). If you do not have a username or password for JPST, you will be required to create an account prior to purchasing.
Full issue PDFs are for PDA members only.
Note to pda.org users
The PDA and PDA bookstore websites (www.pda.org and www.pda.org/bookstore) are separate websites from the PDA JPST website. When you first join PDA, your initial UserID and Password are sent to HighWirePress to create your PDA JPST account. Subsequent UserrID and Password changes required at the PDA websites will not pass on to PDA JPST and vice versa. If you forget your PDA JPST UserID and/or Password, you can request help to retrieve UserID and reset Password below.