Abstract
Recently, the wireless thermologger system was developed. It consists of a wireless logger and the attached computer system. The wireless logger has a wireless structure and can be placed anywhere in the sterilizer chamber for thermometry.
In the present study, thermal evaluations were performed twice using the wireless thermologger system in the sterilizer chamber of the prevacuum autoclave unequipped with sensor ports when it failed to pass the Bowie-Dick test and underwent mechanical repairs. Thermometry was performed when the Bowie-Dick test cycle was operated. The heat up time was measured in the range of 120.0-134.0 °C. The F0 value was calculated with the reference temperature at 121.0 °C and Z value at 10 °C when the sterilization temperature was over 120.0 °C.
The first thermal evaluation was performed after three air-leaking points were repaired along with the replacement of the vacuum pump and the thermosensor. It revealed that the heat-up time was 4 min and 42 s and the F0 value was 137.5. After the temperature control systems were adjusted using the process calibrator in the prevacuum autoclave, the second thermal evaluation revealed that the heat up time was 2 mins 1 s and the F0 value was 102.7.
The present study suggests that the wireless thermologger system is useful in the management of the autoclave unequipped with sensor ports when it undergoes mechanical repairs.
Footnotes
- Copyright © Parenteral Drug Association. All rights reserved.
PDA members receive access to all articles published in the current year and previous volume year. Institutional subscribers received access to all content. Log in below to receive access to this article if you are either of these.
If you are neither or you are a PDA member trying to access an article outside of your membership license, then you must purchase access to this article (below). If you do not have a username or password for JPST, you will be required to create an account prior to purchasing.
Full issue PDFs are for PDA members only.
Note to pda.org users
The PDA and PDA bookstore websites (www.pda.org and www.pda.org/bookstore) are separate websites from the PDA JPST website. When you first join PDA, your initial UserID and Password are sent to HighWirePress to create your PDA JPST account. Subsequent UserrID and Password changes required at the PDA websites will not pass on to PDA JPST and vice versa. If you forget your PDA JPST UserID and/or Password, you can request help to retrieve UserID and reset Password below.