Abstract
The following study was conducted to determine the effect of different preservatives commonly used in the biopharmaceutical industry on the product-specific bubble point of sterilizing-grade filters when used to filter product processed with different types of tubing. The preservatives tested were 0.25% phenol, m-cresol, and benzyl alcohol. The tubing tested was Sani-Pure® (platinum-cured silicone tubing), Versilic™ (peroxide-cured silicone tubing), C-Flex,® Pharmed,® and Cole-Parmer® (BioPharm silicone tubing). The product-specific bubble point values of sterilizing grade filters were measured after the recirculation of product through the filter and tubing of different types of materials for a total contact time of 15 h. When silicone tubing was used, the post-recirculation product-specific bubble point was suppressed on average 13 psig when compared to the pre- recirculation product-specific bubble point. Suppression was also observed with C-Flex,® but to a much lesser extent than with silicone tubing. Suppression was not observed with Pharmed® or BioPharm tubing. Alcohol extractions performed on the filters that experienced suppressed bubble points followed by Fourier transform infrared spectroscopy analysis indicated the filters contained poly(dimethylsiloxane). Direct addition of poly(dimethlysiloxane) to solutions filtered through sterilizing-grade filters suppressed the filter bubble points when tested for integrity. Silicone oils most likely reduced the surface tension of the pores in the membrane, resulting in the ability of air (or nitrogen) to pass more freely through the membrane, causing suppressed bubble point test values. The results of these studies indicate that product-specific bubble point of a filter determined with only product may not reflect the true bubble point for preservative-containing products that are recirculated or contacted with certain tubing for 15 h or greater. In addition, tubing material placed in contact with products containing preservatives should be evaluated for impact to the product-specific bubble point when being utilized with sterilizing-grade filters.
- ATR
- Attenuated total reflectance
- BP
- bubble point
- CPBP
- Corrected product-wetted bubble point
- FTIR
- Fourier transform infrared spectroscopy
- HPLC-UV
- high-performance liquid chromatography—ultraviolet
- IPA
- Isopropyl alcohol
- MMWBP
- Minimum manufacturers water-wetted bubble point specification
- LC/MS
- Liquid chromatography/mass spectroscopy
- PDMS
- Poly(dimethylsiloxane)
- THF
- Tetrahydrofuran
- psig
- Pounds per squareinch gauge
- USP
- United States Pharmacopeia
Footnotes
- Copyright © Parenteral Drug Association. All rights reserved.
PDA members receive access to all articles published in the current year and previous volume year. Institutional subscribers received access to all content. Log in below to receive access to this article if you are either of these.
If you are neither or you are a PDA member trying to access an article outside of your membership license, then you must purchase access to this article (below). If you do not have a username or password for JPST, you will be required to create an account prior to purchasing.
Full issue PDFs are for PDA members only.
Note to pda.org users
The PDA and PDA bookstore websites (www.pda.org and www.pda.org/bookstore) are separate websites from the PDA JPST website. When you first join PDA, your initial UserID and Password are sent to HighWirePress to create your PDA JPST account. Subsequent UserrID and Password changes required at the PDA websites will not pass on to PDA JPST and vice versa. If you forget your PDA JPST UserID and/or Password, you can request help to retrieve UserID and reset Password below.