Abstract
Dilution of samples with water (water dilution) was not appropriate to measure the endotoxin activity in solutions containing chelating agents and detergents, typical formulations showing low endotoxin recovery (LER). Dilution of the samples with 2 mM magnesium solution (magnesium dilution) or addition of the samples directly to the Limulus amebocyte lysate (direct method) provided accurate endotoxin values in the samples. The difference between the water dilution and the magnesium dilution/the direct method seemed to be caused by endotoxin activity decrease during the water dilution. Endotoxin activity was maintained for more than a month in LER solutions at 4oC, when the activity was measured by the magnesium dilution or the direct method. The magnesium concentration in the diluent for the magnesium dilution should be greater than the concentration of the chelating agent in the sample. The magnesium dilution is the most appropriate dilution method for endotoxin measurement in LER solutions. Sodium citrate and polysorbate 20 showed stronger LER effects than phosphate buffer and polysorbate 80, respectively. Human serum albumin was also found to be an LER mitigating factor, and this suggested that protein might reduce LER effects.
- Bacterial Endotoxins Test
- Endotoxin
- Limulus amebocyte lysate
- Lipopolysaccharide
- Low endotoxin recovery (LER)
- Magnesium dilution method
- Received September 5, 2018.
- Accepted March 6, 2019.
- Copyright © 2019, Parenteral Drug Association
PDA members receive access to all articles published in the current year and previous volume year. Institutional subscribers received access to all content. Log in below to receive access to this article if you are either of these.
If you are neither or you are a PDA member trying to access an article outside of your membership license, then you must purchase access to this article (below). If you do not have a username or password for JPST, you will be required to create an account prior to purchasing.
Full issue PDFs are for PDA members only.
Note to pda.org users
The PDA and PDA bookstore websites (www.pda.org and www.pda.org/bookstore) are separate websites from the PDA JPST website. When you first join PDA, your initial UserID and Password are sent to HighWirePress to create your PDA JPST account. Subsequent UserrID and Password changes required at the PDA websites will not pass on to PDA JPST and vice versa. If you forget your PDA JPST UserID and/or Password, you can request help to retrieve UserID and reset Password below.