Abstract
The early detection of microbial contamination is an important issue in the production of biopharmaceuticals using animal cell culture systems. A new method, based on the ChemScan® RDI analyser, was evaluated for the rapid detection and enumeration of low concentrations of microorganisms within a large population of animal cells and in cell culture media. The method was tested with suspension cell cultures and is applicable to adherent cell cultures. In both cases, the method uses an initial step to eliminate the animal cells followed by collection and fluorescent labelling of the viable microbial contaminants on a filter membrane. Total counts of the viable microorganisms were obtained after analysis of the membrane by the ChemScan® RDI analyser. The results showed that the ChemScan® RDI detected individual bacterial cells after filtration of the pre-treated animal cell culture. The detection limit of the ChemScan® RDI was less than 10 bacteria/ml in cell culture containing 106 mammalian cells/ml and one bacterium in 500 ml of cell culture medium. A strong correlation between the standard plate count and the ChemScan® RDI was observed, even at low bacterial concentrations. The total time for each analysis was less than two hours.
Footnotes
- Copyright © Parenteral Drug Association. All rights reserved.
PDA members receive access to all articles published in the current year and previous volume year. Institutional subscribers received access to all content. Log in below to receive access to this article if you are either of these.
If you are neither or you are a PDA member trying to access an article outside of your membership license, then you must purchase access to this article (below). If you do not have a username or password for JPST, you will be required to create an account prior to purchasing.
Full issue PDFs are for PDA members only.
Note to pda.org users
The PDA and PDA bookstore websites (www.pda.org and www.pda.org/bookstore) are separate websites from the PDA JPST website. When you first join PDA, your initial UserID and Password are sent to HighWirePress to create your PDA JPST account. Subsequent UserrID and Password changes required at the PDA websites will not pass on to PDA JPST and vice versa. If you forget your PDA JPST UserID and/or Password, you can request help to retrieve UserID and reset Password below.